A gas sensor gives different resistance readings at two humidity levels. Does that mean the sensor is sensitive to humidity? Before drawing that conclusion, check whether the same concentration of target gas reached the sensor surface in both runs. Adding water vapor changes both the relative humidity and the fraction of target gas in the mixture. The concentration setpoint on the gas blender alone cannot tell you which change affected the reading.
Tianjin University of Technology specified three uses for its dynamic gas blending system with temperature and humidity control: testing gas sensor performance, simulating changes in gas composition and humidity together, and calibrating sensors for research. The contract award notice names Beijing Airppb Environmental Protection Equipment Co., Ltd. as the supplier and identifies the model as the Airppb MR-D02. For this project, Airppb designed a system that controls humidity by blending dry and humidified carrier gas. A water vapor generator humidifies one stream, and the system adjusts the ratio of the two streams using feedback from a calibrated temperature and humidity probe. A separate port in the calibration chamber accepts an independent probe so researchers can check conditions near the sensor itself. The MR-D02 product literature also describes metering liquid with a syringe pump and then vaporizing it. The humidity control method specified for this project uses the dry and humidified carrier gas streams described here.
“Test at 100 ppm at both humidity levels” can mean two things. The dry gas may be prepared at 100 ppm before water is added in each run: the dry-basis concentration stays constant. Or the humidified mixture reaching the sensor may need to contain 100 ppm in each run: the wet-basis concentration stays constant. These two conditions call for different gas mixtures.
Suppose the dry gas contains 100 ppm of the target species. If humidification adds only water vapor, none of the target species is lost, and water accounts for 2% of the total moles in the final mixture, the wet-basis concentration is 100 × (1 − 0.02) = 98 ppm. Here, 2% is the mole fraction of water vapor. The relative humidity reading (%RH) cannot be substituted for it. This is a worked example, not a measurement from the project.
The project design assigns a separate mass flow controller (MFC) to each of six target-gas channels. The researcher specifies the source gas concentrations, target composition and total flow. The software allocates flow to each channel and checks that the required flow falls within its effective control range. To hold wet-basis concentration constant, the calculation must include water vapor in the total. To establish that the actual concentrations at the two humidity levels are comparable, use a reference method suited to the target species and humidity, at the agreed temperature, pressure and measurement location.
Before entering the water vapor generator, dry carrier gas passes through either the main or the auxiliary flow controller. The two branches are interlocked: only one operates at a time. Consider a gas mixture that requires 8 sccm of dry carrier gas at the generator inlet.
| Humidification branch | Full-scale flow | 8 sccm as a share of full scale | With an effective lower limit of 2% of full scale |
| Main | 1,000 sccm (1 slm) | 0.8% | Below the effective range |
| Auxiliary | 50 sccm | 16% | Within the effective range |
At 8 sccm, selecting the auxiliary branch puts the controller within its effective operating range. The 8 sccm value is a carrier gas flow rate; it does not correspond to a fixed relative humidity. After switching ranges, wait until the temperature and humidity readings again meet the agreed stability criteria before comparing sensor readings.
One question is easy to miss when requesting a quotation: who will measure the sensor’s electrical signal? The project design includes four types of adapters: for ceramic-tube sensors, four-corner planar devices, micro/nano chips and room-temperature devices with interdigitated electrodes. Airppb supplies the fixtures and wiring adapters; the purchaser supplies the source measure unit (SMU) or electrical data acquisition equipment. For a similar experiment, include the sensor dimensions, pinout and model of your existing acquisition equipment along with the gas species and concentrations.
If you are selecting a dynamic gas blending system for sensor tests, send Airppb’s technical team the gas species and source concentrations, required wet- or dry-basis target concentrations, total flow, humidity levels, exposure sequence and timing, and sensor interface for one experimental run. With that information, the team can assess the flow range for each channel, where to place the reference probe and which quantities to verify at the sensor.
Sources: Public tender notice and tender documents on the China Government Procurement Network; republication of the contract award notice, project no. 0615-2641061250922; and the project technical design and configuration document and accompanying product technical and operating manual from Beijing Airppb Environmental Protection Equipment Co., Ltd. The configuration described is specific to the Tianjin University of Technology project. Acceptance of an individual system depends on its own measured performance.