Day-to-day handling, sampling practice and care of the composite hazardous substance identifier
This guide summarises good field practice for the MR-AIMS. It supports training and pre-shift briefings. Where the official user manual or your standing orders differ, they take precedence.
This guide is for responders, inspectors and technicians who carry the MR-AIMS on duty. It assumes no prior knowledge of ion mobility spectrometry.
Its aim is simple: get a trustworthy result fast, recognise when a result needs caution, and keep the instrument serviceable between services.
Switch on in clean air. Avoid exhausts, solvents and smoking areas. Warm-up takes twelve minutes or less. The drift tube, electronics and sampling path stabilise during this time.
Starting in contaminated air biases the baseline. A clean start-up point is worth the walk. Once warm-up ends, confirm a stable baseline before moving in.
If the unit reports a fault at start-up, do not deploy it. Record the condition and withdraw it for service.
The unit reports the best library match and the substance class. Treat it as decision support, not absolute truth. Ask whether the result fits the location.
A high-toxicity compound in a paint store deserves scepticism. Hydrogen sulphide at a wastewater inlet does not. Mixed atmospheres are the hardest case.
One substance can dominate and suppress another. Sample from several positions. Record every identification, not only the first alarm.
Before switching off, run the unit in clean air until the reading clears. This purges the sampling path and reduces carry-over into the next job.
Wipe the exterior. Decontaminate if exposure is suspected. Charge the battery. Complete the log. Report anomalies while details are fresh.
| Interval | Action |
|---|---|
| After each use | Purge in clean air, wipe, charge, update log. |
| Daily (in service) | Check inlet and filter, charge, library and self-test status. |
| Weekly | Inspect seals, ports and connectors; clean; review fault messages. |
| Monthly | Functional check in clean air and against a verification standard. |
| Annually (or per contract) | Manufacturer service, verification and library update. |
| Symptom | Likely Cause | Action |
|---|---|---|
| Will not power on | Battery flat or unseated | Refit or replace; charge before reuse. |
| Warm-up stalls | Extreme temperature or fault | Move inside −20 °C to +45 °C; else withdraw. |
| Reading lingers in clean air | Contaminated inlet | Purge longer; clean or replace consumables. |
| Repeated odd alarms | Solvents, fuels, humidity | Clear in clean air; re-sample; note interference. |
| Short runtime | Aged battery or cold | Carry a spare; expect ≥ 6 h in cold. |
| No identification | Substance not in library | Log it; request a library extension. |
A full charge gives at least ten hours at normal temperature. Cold reduces capacity. Plan for at least six hours in low temperatures.
Keep a spare pack warm inside clothing during winter work. Top up whenever you can. Do not wait for the low-battery warning.
Store between −40 °C and +55 °C. Keep the unit clean, dry and protected from impact, ideally in its case.
For long storage, remove or periodically recharge the battery. Cap or cover the inlet. The IP67 rating tolerates wet and dusty service, but dry the unit before storage.
For quotations, demonstrations, library customisation and technical support, contact the manufacturer directly.
| Channel | Details |
|---|---|
| Company | Beijing Airppb Environmental Protection Equipment Co., Ltd. |
| Address | 4th Floor, Building 1, No. 11 Kechuang 14th Street, Beijing Economic-Technological Development Area, Beijing, China |
| Telephone / Fax | Tel: +86-10-50928608 | Fax: +86-10-56865313 | WhatsApp: +86-13811100776 |